You are here: Home / Products / Hand Held Digital Instruments
Our Highest Precision Hand-Held Temp/Humidity (CO2 with optional sensor) Meter with Watertight Case, 8130.TFF Sensor, Cable and Batteries!
Model: 5810-20
NIST Certification available.
Highest-precision measurements for industrial temperature and humidity calibrations The XP-201 (Model 5810-20) is the perfect choice when highest-precision measurements are needed for industrial temperature and humidity calibrations. It is suitable as a humidity reference in climate chambers or humidity generators. A mini USB interface with software and online data collection are included.
The most accurate handheld device (0.5%) for relative humidity. Ideal as a reference standard. Excellent stability, very good repeatable measurements, absolute hysteresis latitude as well as no drift of the measurement value at very high humidity levels. For traceability to national standards a DAkkS calibration certificate is attached. Special features: 0.5% accuracy, high quality watertight case, resistive-electrolytic sensor, 2m cable, on/off switch, Touch Screen, batteries and DAkks calibration certificate are included.
Two page brochure here. See optional 8120-TFF (Temp/RH) and 7120-CO2 (CO2) sensors for use with this meter.
| Sensor power supply | 5.5 V ± 10 % DC, max. 200 mA |
| Integrated air pressure sensor | |
| Measuring range | 800...1100 mbar |
| Accuracy at 25°C, 1013.25mbar | 0.5 mbar |
| Long-term stability | typ. 1mbar/year |
| Measurement resolution | 0.024 mbar |
| Measuring principle | Piezoresistive |
| Calculated measurement categories for external temperature/humidity sensors | Mathematical: MIN/MAX/AVG/HOLD Temperature °C/°F Rel. humidity %r.h Rel. humidity of ice %r.h Water vapour density (absolute humidity) g/m3 Dew point temperature °C/°F Frost point temperature °C/°F Mixing ratio at saturation (100%) g/kg Volume fraction of water vapour /mass fraction of water vapour % Wet-bulb temperature °C/°F Ice-bulb temperature °C/°F Specifi c Enthalpy (mass of air) kJ/kg Saturation vapour pressure above ice/water hPa Vapour particle pressure hPa Air density kg/m3 |
| General | |
| Dimensions | 6.7 X 2.5 X 1.34", 170 x 62 x 34 mm |
| Weight | Approx. 7 1/4oz, 205 g |
| Housing material sensor | PVDF black |
| Mechanical sensor protection sensor | Standard polyethylene dust filter |
| Storage conditions | |
| Permissible ambient temperature | -10...60 °C |
| Operating rel. humidity | <90 % RH non - condensing |
| Operating conditions | |
| Operating rel. humidity | <90 % RH non - condensing |
| Operating temperature | -20...80 °C |
| Admissible height above absolute altitude | 4000 m |
| Power supply | |
| Power supply | 4 Alkaline LR6 AA 1.5 V/USB 5 V |
| Active power consumption | Approx. 400 mW |
| Battery life passive | Approx. 1 year |
| Battery life active | min. 24 hours |
| Data storage | |
| Integrated data storage | Up to 200 gauges taking approx. 1 mill. values |
| Interface | |
| USB | Cable and SmartGraph3 software included in delivery |
| Display | |
| Definition of measured values | 2 decimal places |
| Display | |
| Control | Touch screen, capacitive |
| Technology | TFT, resolution 240 x 320, 65 k colours, very good contrast due to Piezoresistive technology |
| Surface, toughened glass | Degree of hardness: 7, scratch - resistant |
| Relative Humidity | |
| Principle | Resistive - electrolytic |
| Measuring range | 0 ... 100 % |
| Unit | % |
| Measurement accuracy incl. reproducibility and hysteresis | :Humidity*: 15...30 °C, ±0.5 % RH | 0...50 °C, ±0.8 % RH | -20...80 °C, ±2.5 % RH (*The humidity accuracy refers to the nominal values of Novasina humidity standards, which refer to the Greenspan Report.) |
| Number of calibration points | :Humidity: 13 points over full measurement range | Temperature: 2 points over full measurement range |
| Temperature | |
| Principle | NTC |
| Measuring range | -20 ... 80 °C |
| Unit | °C |
| Accuracy | 0.15 °C between 0 +70 °C, otherwise 0.25 °C |